Document #:
Beams-doc-722-v1
Document type:
Note
Submitted by:
Alvin Tollestrup
Updated by:
Alvin Tollestrup
Document Created:
08 Jul 2003, 22:26
Contents Revised:
08 Jul 2003, 22:26
Metadata Revised:
08 Jul 2003, 22:26
Viewable by:
  • Public document
Modifiable by:
  • Same as Viewable by

Quick Links:
Latest Version

Abstract:
This note describes the procedure used to calibrate the SBD using T:IBEAM as a reference. We give the calibration for Stores 2495, 2502, 2503, 2505 and 2507. The fluctuations between stores are a few tenths of a percent. There is also a measure of the DC beam accumulated at injection which seems to be 1-3 percent of the T:IBEAM. The calibration can be checked on line by comparing T:SBDTWG (the SBD sum of p and pbar) with T:IBEAM. Some information on the statistical noise is given. The statistical fluctuation of SBDPIS / AIS is measured to be about 0.3% and data is also given for SBDPWS / AWS.
Files in Document:
Keywords:
SBD cable dispersion